BeOSL Photon Dosimeter

Photon dosimeters are used, among other things, by employees who perform work with X-ray equipment, such as in hospitals, veterinary clinics and dental practices or where there are gamma radiation emitting sources.

The BeOSL photon dosimeter is equipped with two detectors of beryllium oxide. These are located behind different filters of 54 and 616 mg/cm2 respectively. The first detector is used to determine the surface dose, the second to quantify the depth dose.

Detectors: Beryllium oxide
Mounting options: Various, including via a clip
Sterilizability: Must not be sterilized
Disinfectability: Alcohol consumption is permitted
Change dosimeters: Monthly
Measured variable: HP(10) en HP(0.07) in mSv
Photon detection limits: 0.05 mSv
Maximum photon dose range: 1 Sv
Energy range photons: 30 keV to 1.3 MeV
Energy and angle dependence: See image below

Explanation of personal dosimetry >>